PRODUCT SHEET

Combined Holmium/Neodymium Oxide Solution (220 – 865 nm)

This reference material, known as HNL, is formulated for wavelength scale qualification of UV-visible spectrophotometers operating within the 240 to 795 nm spectral range. It is suitable for instruments with a spectral bandwidth of 5 nm or less.

The solution consists of Holmium oxide and Neodymium oxide dissolved in 10% v/v perchloric acid, permanently sealed by heat fusion inside a 10 mm far UV quartz cuvette. This composition provides stable and interference-free absorption peaks across a broad wavelength span, improving over previous references by extending the valid certified peak range from 219 nm to 864 nm.

Features and Purposes
  • Contains 14 certified absorption peaks at nominal wavelengths: 219, 241, 287, 361, 416, 485, 522, 537, 575, 641, 732, 740, 794, and 864 nm.
  • Peak values are spectral bandwidth dependent due to asymmetric absorption bands; certified values are provided at bandwidths of 0.10, 0.25, 0.50, 1.00, 1.50, 2.00, 3.00, 4.00, and 5.00 nm.
  • Supplied in a 10 mm pathlength far UV quartz cuvette, permanently sealed for long-term stability.
  • Certified calibration certificates accompany each unit, with the option to request certified values at additional wavelengths and bandwidths.
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy
Starna Cell Catalogue Micro Rectangular Spectrophotometer Cells
Combined Holmium Neodymium Oxide Liquid Reference for UV and Visible Spectrophotometer Wavelength Qualification
PRODUCT SHEET

Combined Holmium/Neodymium Oxide Solution (220 – 865 nm)

This reference material, known as HNL, is formulated for wavelength scale qualification of UV-visible spectrophotometers operating within the 240 to 795 nm spectral range. It is suitable for instruments with a spectral bandwidth of 5 nm or less.

The solution consists of Holmium oxide and Neodymium oxide dissolved in 10% v/v perchloric acid, permanently sealed by heat fusion inside a 10 mm far UV quartz cuvette. This composition provides stable and interference-free absorption peaks across a broad wavelength span, improving over previous references by extending the valid certified peak range from 219 nm to 864 nm.

Features and Purposes
  • Contains 14 certified absorption peaks at nominal wavelengths: 219, 241, 287, 361, 416, 485, 522, 537, 575, 641, 732, 740, 794, and 864 nm.
  • Peak values are spectral bandwidth dependent due to asymmetric absorption bands; certified values are provided at bandwidths of 0.10, 0.25, 0.50, 1.00, 1.50, 2.00, 3.00, 4.00, and 5.00 nm.
  • Supplied in a 10 mm pathlength far UV quartz cuvette, permanently sealed for long-term stability.
  • Certified calibration certificates accompany each unit, with the option to request certified values at additional wavelengths and bandwidths.
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy
Starna Cell Catalogue Micro Rectangular Spectrophotometer Cells
Combined Holmium Neodymium Oxide Liquid Reference for UV and Visible Spectrophotometer Wavelength Qualification

Contact us to discover the offer dedicated to this product!

Compila il form per richiedere un preventivo.

Compila il form per richiedere un preventivo.