PRODUCT SHEET

Quinine Sulfate Solution References

This Reference Material consists of 10 mm far UV quartz fluorometer cuvettes with polished faces at 90°, containing Quinine Sulfate dissolved in perchloric acid and permanently sealed by heat fusion. It is supplied either with or without a perchloric acid blank, and is also available as sets of solutions with concentrations up to 1.0 mg/l for use as a linearity check.

Quinine Sulfate Dihydrate in perchloric acid is an established reference material in fluorescence spectroscopy, designated SRM 936a by NIST. The provided data are based on its corrected fluorescence emission spectrum with an excitation wavelength of 347.5 nm, enabling reliable calibration of spectrofluorometers over the range 375 nm to 675 nm.

Features and Purposes
  • Calibrates spectral response of spectrofluorometers from 375 nm to 675 nm
  • Available in single cuvette or sets with multiple concentrations: 0.25, 0.5, 0.75, and 1 mg/l
  • Cuvettes made of far UV quartz, 10 mm pathlength, polished faces at 90°
  • Sealed by heat fusion to ensure sample integrity
  • Includes perchloric acid solvent with optional blank cell
  • Can be used for linearity checks within low concentration range (up to 1 mg/l)
  • Certified fluorescence emission data provided in calibration certificate at 5 nm intervals
  • Allows calculation of correction factors for instrument spectral response and corrected emission intensity
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy
Starna Cell Catalogue Original Micro Rectangular Cells with Lid and Stopper
Quinine Sulfate Solution Spectrofluorometer Calibration References
PRODUCT SHEET

Quinine Sulfate Solution References

This Reference Material consists of 10 mm far UV quartz fluorometer cuvettes with polished faces at 90°, containing Quinine Sulfate dissolved in perchloric acid and permanently sealed by heat fusion. It is supplied either with or without a perchloric acid blank, and is also available as sets of solutions with concentrations up to 1.0 mg/l for use as a linearity check.

Quinine Sulfate Dihydrate in perchloric acid is an established reference material in fluorescence spectroscopy, designated SRM 936a by NIST. The provided data are based on its corrected fluorescence emission spectrum with an excitation wavelength of 347.5 nm, enabling reliable calibration of spectrofluorometers over the range 375 nm to 675 nm.

Features and Purposes
  • Calibrates spectral response of spectrofluorometers from 375 nm to 675 nm
  • Available in single cuvette or sets with multiple concentrations: 0.25, 0.5, 0.75, and 1 mg/l
  • Cuvettes made of far UV quartz, 10 mm pathlength, polished faces at 90°
  • Sealed by heat fusion to ensure sample integrity
  • Includes perchloric acid solvent with optional blank cell
  • Can be used for linearity checks within low concentration range (up to 1 mg/l)
  • Certified fluorescence emission data provided in calibration certificate at 5 nm intervals
  • Allows calculation of correction factors for instrument spectral response and corrected emission intensity
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy
Starna Cell Catalogue Original Micro Rectangular Cells with Lid and Stopper
Quinine Sulfate Solution Spectrofluorometer Calibration References

Contact us to discover the offer dedicated to this product!

Compila il form per richiedere un preventivo.

Compila il form per richiedere un preventivo.