PRODUCT SHEET

High Purity Water Sensitivity Reference

This product is a 10mm far UV quartz fluorometer cuvette with polished faces at 90°, filled with ultra-high purity water and permanently sealed by heat fusion. It serves as a reference material for evaluating the signal-to-noise ratio (SNR) of a spectrofluorometer near its detection limit, effectively indicating the instrument’s ultimate sensitivity.

Unlike traditional methods using dilute fluorophore solutions, which are challenging and prone to contamination at low concentrations, this reference employs the phenomenon of Raman scattering. When excited by UV or visible radiation, pure water produces a characteristic Raman peak due to Stokes Raman scattering at a lower energy (longer wavelength). The resulting signal levels are exceptionally low and comparable to those from trace analyte concentrations, making it ideal for sensitivity qualification.

Features and Purposes
  • 10mm pathlength far UV grade quartz cuvette with polished 90° faces
  • Contains ultra-high purity water, permanently sealed by heat fusion
  • Utilizes Raman scattering phenomenon for signal-to-noise ratio evaluation
  • Enables testing near the spectrofluorometer detection limit without sample preparation errors
  • Suitable excitation wavelengths include 350nm and 500nm for typical spectral evaluation
  • Catalog number: RM-H2O
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy 2024
Starna Cell Catalogue Micro Rectangular Spectrophotometer Cells
High Purity Water Raman Reference for Spectrofluorometer Sensitivity and Signal to Noise Ratio Evaluation
PRODUCT SHEET

High Purity Water Sensitivity Reference

This product is a 10mm far UV quartz fluorometer cuvette with polished faces at 90°, filled with ultra-high purity water and permanently sealed by heat fusion. It serves as a reference material for evaluating the signal-to-noise ratio (SNR) of a spectrofluorometer near its detection limit, effectively indicating the instrument’s ultimate sensitivity.

Unlike traditional methods using dilute fluorophore solutions, which are challenging and prone to contamination at low concentrations, this reference employs the phenomenon of Raman scattering. When excited by UV or visible radiation, pure water produces a characteristic Raman peak due to Stokes Raman scattering at a lower energy (longer wavelength). The resulting signal levels are exceptionally low and comparable to those from trace analyte concentrations, making it ideal for sensitivity qualification.

Features and Purposes
  • 10mm pathlength far UV grade quartz cuvette with polished 90° faces
  • Contains ultra-high purity water, permanently sealed by heat fusion
  • Utilizes Raman scattering phenomenon for signal-to-noise ratio evaluation
  • Enables testing near the spectrofluorometer detection limit without sample preparation errors
  • Suitable excitation wavelengths include 350nm and 500nm for typical spectral evaluation
  • Catalog number: RM-H2O
DOCUMENTS AND RESOURCES
Certified Reference Materials for Spectroscopy 2024
Starna Cell Catalogue Micro Rectangular Spectrophotometer Cells
High Purity Water Raman Reference for Spectrofluorometer Sensitivity and Signal to Noise Ratio Evaluation

Contact us to discover the offer dedicated to this product!